• Nu esti logat

    Intra in cont

FlexTop - 2221


Senzori temperatura


Producător: Baumer Disponibilitate: Solicita stoc Cantitate disponibila: 0 buc.
Solicita produs
  • For RTD (resistance) or C/T thermocouple)
  • Input: RTD, T/C, mV and ohm
  • Measuring range -270 ... +2300 °C
  • More accurate than 0.1 °C for Pt100
  • ATEX explosion protection
  • DIN form B housing installation
  • HART protocol

Output signal

Characteristic:
Linear or customised with max. 30 points

Damping:
0.0 … 15.0 s , programmable

Output signal:
20 … 4 mA , 2-wire + HART®
4 … 20 mA , 2-wire + HART®

Accuracy:
< 0.1 % FSR

Temperature drift:
< 0.003 %/K , typ.
< 0.01 %/K , max.

Load resistance:
R = (U - 8 V)/23 mA

Response time, T90:
≤ 1.0 s , with Pt100
≤ 1.6 s , with T/C

Resolution:
12 bit

Up/Down scaling limits:
23 mA / 3.5 mA

Ripple immunity:
3 Vrms

Power-up time:
1.8 … 3.9 s

Power supply:
8 … 35 V DC

ambient conditions

Degree of protection (EN 60529):
IP 10 , terminals
IP 55 , housing

Long-term test (transmitter):
IEC 770 6.3.2

Humidity:
< 98 % RH , condensing

Compliance and approvals

EMC:
EN 61326-1

Namur:
NE21

Explosion protection:
ATEX II 1G Ex ia IIC T5/T6
ATEX II 3G Ex nA IIC T5

HART® interface

Properties:
2-point current-trim
2-point sensor-trim
Input signal logging
Read input signal value
Read output signal value
Read serial number
Read/Change configuration
Read/Change user ID

Protocol:
HCF standard, Rev.5

Data sheet

Data sheet – FlexTop 2221 - DE Descarca
Data sheet – FlexTop 2221 - EN Descarca
Data sheet – FlexTop 2221 - FR Descarca

Installation and operating instructions

Operating instructions – FlexTop 2221 - EN Descarca

Certificates / approvals

Declaration of Conformity – CE - EN/DE/FR Descarca
Declaration of Conformity – CE - ATEX - EN/DE/FR Descarca
Type Examination Certificate / Type approval – ATEX - EN Descarca

Technical Reports / White Paper

Baumer Guideline for Analog Sensor Interfaces - DE Descarca
Baumer Guideline for Analog Sensor Interfaces - EN Descarca
Baumer Explosion Protection Guideline - DE Descarca
Baumer Explosion Protection Guideline - EN Descarca

Software

Software – DDL - Descarca
Software – PC Software FlexProgram - DE/EN/FR Descarca
Software – FlexProgram HART configuration package - Descarca
Acorda o nota: